Improvement of spatial resolution of µ‐XRF by using a thin metal filter

Kazuhiko Nakano, Akinori Matsuda, Yuta Nodera, Kouichi Tsuji

X-Ray Spectrometry · 2008 · 13 citations · 13 references

Concepts

Abstract

Abstract An improvement of spatial resolution of µ‐XRF by using a thin metal filter was investigated. The size of the x‐ray beam focused by the polycapillary x‐ray lens depended on the energy of the characteristic x‐rays. Original spot sizes at the focal point were 48 µm for CrKα, 41 µm for NiKα, and 28 µm for MoKα, respectively. To make the x‐ray beam size small, TiCu thin foil was placed between the output of the lens and the focal point as a metal filter to reduce the continuous x‐rays. Finally, the x‐ray microbeam size was improved to 30 µm by applying a filter. Clear 2D mapping images of Cr, Fe, and Ni in 300‐mesh stainless steel could be obtained by applying this filter. Copyright © 2008 John Wiley & Sons, Ltd.