X-Ray Spectrometry · 2008 · 13 citations · 13 references
Natural SciencesMaterials SciencePhysicsDiagnostic ImagingMetallurgyOpticsRadiographic ImagingX-ray OpticsX-ray SpectroscopyBeam Spot SizeFiltering Of X-raysSpatial Resolution AnalysisSpatial Resolution MeasurementX-ray Focusing OpticsX-ray Focusing SystemsX-ray Imaging OpticsX-ray Imaging Physics
Abstract An improvement of spatial resolution of µ‐XRF by using a thin metal filter was investigated. The size of the x‐ray beam focused by the polycapillary x‐ray lens depended on the energy of the characteristic x‐rays. Original spot sizes at the focal point were 48 µm for CrKα, 41 µm for NiKα, and 28 µm for MoKα, respectively. To make the x‐ray beam size small, TiCu thin foil was placed between the output of the lens and the focal point as a metal filter to reduce the continuous x‐rays. Finally, the x‐ray microbeam size was improved to 30 µm by applying a filter. Clear 2D mapping images of Cr, Fe, and Ni in 300‐mesh stainless steel could be obtained by applying this filter. Copyright © 2008 John Wiley & Sons, Ltd.