Loss of Tumor Marker-Immunostaining Intensity on Stored Paraffin Slides of Breast Cancer

295

Citations

1

References

1996

Year

Abstract

Slide storage, particularly at room temperature, results in substantial loss of p53 reactivity, with some p53-positive cases becoming p53 negative after 12 weeks of storage. Substantial loss of immunoreactivity for factor VIII, ER, and Bcl-2 occurs on slides stored at room temperature for 12 weeks. Storage of unstained slides for up to 12 weeks may lead to false-negative immunostaining for p53 and other antigens.