Radiology · 1987 · 93 citations · 12 references
Theoretical AnalysisEngineeringMedical ImagingPet-mriAdvanced ImagingBiomedical ImagingMagnetic ResonanceNeuroimagingInverse ProblemsNeurologyMri-guided Radiation TherapyPartial Flip SequencesMedical Image ComputingMedicineMagnetic Resonance ImagingRadiology
Theoretical analysis predicts that performing magnetic resonance (MR) imaging with partial (less than 90 degrees) flip angles can reduce imaging times two- to fourfold when lesions with elevated T1 values are being examined. This time savings occurs because repetition time (TR) is reduced when imaging is performed with partial flips. Partial flip MR imaging can also improve signal-to-noise ratio (S/N) in fast body imaging. For this study, analytical tools were used to predict image contrast and S/N for short TR, partial flip sequences. Experimental implementation of the short TR, partial flip sequences that analytical work had predicted would be optimal supported the analytical predictions and demonstrated their validity. Partial flip MR imaging is applicable to reducing imaging time only when the ratio of signal differences to noise exceeds threshold values in conventional MR images. Partial flip sequences can be used to advantage in MR imaging of both the head and the body, and the observed effects are predictable through theoretical analysis.
12
Halving MR imaging time by conjugation: demonstration at 3.5 kG.
David A. Feinberg, J D Hale, J C Watts et al. · Radiology · 1986 · 259 citations
Nuclear magnetic resonance whole-body imager operating at 3.5 KGauss.
L E Crooks, M Arakawa, J Hoenninger et al. · Radiology · 1982 · 237 citations
Nuclear magnetic resonance imaging in multiple sclerosis
Steven A. Lukes, Lawrence E. Crooks, Michael J. Aminoff et al. · Annals of Neurology · 1983 · 200 citations
Magnetic resonance of the brain: the optimal screening technique.
Michael Brant‐Zawadzki, David Norman, T H Newton et al. · Radiology · 1984 · 160 citations