Optics Express · 2011 · 43 citations · 16 references
EngineeringBioengineeringBiomedical EngineeringBiomedical OpticsBiophotonicsSpectroscopyInfrared SpectroscopyLight-tissue InteractionNear-infrared ImagingNear-infrared SpectroscopyOptical ImagingOptical SpectroscopyDetector Spatial ResolutionDiffuse Optical ImagingDiffuse Reflectance ImagingDiffuse Reflectance SpectroscopyNon-contact MeasurementNon-contact SensingSpatial Resolution MeasurementTime-domain Near-infrared Spectroscopy
We report results of the proof-of-principle tests of a novel non-contact tissue imaging system. The system utilizes a quasi-null source-detector separation approach for time-domain near-infrared spectroscopy, taking advantage of an innovative state-of-the-art fast-gated single photon counting detector. Measurements on phantoms demonstrate the feasibility of the non-contact approach for the detection of optically absorbing perturbations buried up to a few centimeters beneath the surface of a tissue-like turbid medium. The measured depth sensitivity and spatial resolution of the new system are close to the values predicted by Monte Carlo simulations for the inhomogeneous medium and an ideal fast-gated detector, thus proving the feasibility of the non-contact approach for high density diffuse reflectance measurements on tissue. Potential applications of the system are also discussed.
16
Alwin Kienle, Lothar Lilge, Michael S. Patterson et al. · Applied Optics · 1996 · 498 citations
Adam Liebert, Heidrun Wabnitz, Jens Steinbrink et al. · Applied Optics · 2004 · 268 citations