A Differential Refractometer*

B. Ba

Journal of the Optical Society of America · 1951 · 200 citations · 10 references

Concepts

TL;DR

The paper presents the design and performance of an absolute‑deviation differential refractometer. The instrument employs a square differential cell with a thin partition, mercury and sodium lamps, monochromatic filters, a slit, a projection lens, and a micrometer microscope, all precisely aligned on an optical bench, and can be rotated 180° to swap solvent and solution, effectively doubling the deviation. Ray tracing shows that the slit‑image separation Δd is accurately proportional to the refractive‑index difference Δn, yielding a 0.01‑unit range, 3 × 10⁻⁶ sensitivity, 0.5 % accuracy, and making the device suitable for measuring small refractive‑index or concentration differences and for determining refractive‑index increments in light‑scattering molecular‑weight studies.

Abstract

The design and performance of an absolute deviation type of differential refractometer is described. The apparatus comprises mercury and sodium lamps, monochromatic filters, slit, differential cell, projection lens, and micrometer microscope, all accurately aligned on an optical bench. The cell is square with a thin partition separating solution and solvent, such that the angle of incidence is about 69° on the interface. The cell and its holder, in a jacketed housing, can be turned through 180° in order to interchange solvent and solution and approximately double the deviation. By ray tracing it is shown that the distance between the slit images in the eyepiece field, Δd, is accurately proportional to the difference in refractive index, Δn, of solution and solvent, and that the factor of proportionality can be evaluated from accurately measurable geometrical quantities: the cell partition angle, the virtual distance from slit to cell center, and the magnification of the system up to the field of the micrometer eyepiece. The range of the instrument is approximately 0.01 unit and the limiting sensitivity about 3 × 10−6 unit of refractive index difference. The accuracy in determination of Δn is about 0.5 percent. The instrument is useful for determination of small differences in refractive index or concentration, and for determination of refractive index increments needed in the light-scattering method of evaluating high molecular weights.

References

10