Optics Express · 2012 · 30 citations · 33 references
EngineeringBioengineeringBiomedical EngineeringBiomedical ImagingBiomedical OpticsDiagnostic ImagingMedical ImagingOptical CoherenceOptical Coherence TomographyOptical ImagingOptical TomographyTissue OpticsOptical Coherence Tomography InstrumentationOptical Coherence Tomography Tissue PropertiesPolarization PropertiesSingle-camera ActingSpectral-domain Optical Coherence TomographySystem Calibration
Jones matrix optical coherence tomography can fully characterize depth-resolved polarization properties in tissue. In this report, we described a simple single-camera based implementation of full-range spectral domain Jones matrix optical coherence tomography. The Jones matrix reconstruction algorithm was described in detail and system calibration was demonstrated with comprehensive examples. In addition to the conventional structural image, the images of retardance, optical axis and relative attenuation can be obtained from the measured Jones matrix image. Both in vitro and in vivo image examples were presented to demonstrate the polarization imaging ability of the system.
33
David Huang, Eric A. Swanson, Charles P. Lin et al. · Science · 1991 · 13.7K citations
Christoph K. Hitzenberger, Erich Goetzinger, Markus Sticker et al. · Optics Express · 2001 · 391 citations · Full text