2009 · 41 citations · 19 references
EngineeringBiomedical EngineeringElectrical EngineeringBioinstrumentationBiomemsMicrowave EngineeringImplant DesignImplant TechnologyImplantable Biomedical DevicesImplantable MicrodevicesMedical Device DesignWireless Power TransferBiomedical ImplantsCoil DesignDesign MethodologyDesign PrinciplesImplantable Wireless Power TransferInductive Power TransferPower OutputWireless Power Transfer For Implants
This paper describes design principles to design and develop a transcutaneous link for medical implants using inductively coupled coils. Parameters which optimize link efficiency have been discussed in the light of previous studies and a simple design methodology to find optimized parameters for Class E amplifier and inductive coils is outlined. Paper also describes design of an indigenously developed transcutaneous link from commercial off-the-shelf components to demonstrate the design process. Simulation and practical results of the link developed at 2.5MHz for 100mW output power are provided. We were able to achieve 40% link efficiency with data rate of 128kbps from laboratory-based discrete electronic components.
19