Variability in Effective Radiating Area at 1 MHz Affects Ultrasound Treatment Intensity

Stephen J. Straub, Lennart D. Johns, Samuel M. Howard

Physical Therapy · 2007 · 30 citations · 22 references

DOIFull text

Open access

Concepts

Abstract

BACKGROUND AND PURPOSE: Previous research has indicated that not all ultrasound transducers heat at equal rates; however, the cause of this disparity is unclear. Variability in spatial average intensity (SAI) has been implicated in this disparity at 3 MHz. This variability has not been explored at 1 MHz. METHODS: Sixty-six 5-cm(2) ultrasound transducers were purchased from 6 different manufacturers. Transducers were calibrated and assessed for effective radiating area (ERA), total output power, and SAI using standardized measurement techniques. RESULTS: Total output power values fell within US Food and Drug Administration guidelines, but there were large variations in ERA. The resulting SAI values showed large deviations (-43% to +61%) from the digitally displayed value. Intra-manufacturer SAI values varied up to 53%. DISCUSSION AND CONCLUSION: Spatial average intensity can vary largely from the values displayed on these ultrasound generators; in a calibrated cohort, this difference is primarily attributable to differences in measured ERA. Patterns of SAI variability within the manufacturer at 1 MHz do not follow previous reports of variability at 3 MHz.

References

22