British Journal of Radiology · 1992 · 83 citations · 9 references
MedicineRadiologyDiagnostic ImagingDiagnostic RadiologyOrthopedicsBody CompositionBone HealthBone Quantity AssessmentHuman OsteologyImaging AssessmentMusculoskeletal ImagingBone Density ComparisonBone Density EstimationBone Measurement StandardsBone Mineral DensityRadiographic Bone MeasurementReference Data ComparisonReference Data ValidationStandard Reference Databases
We have investigated whether the Lunar DPX (software 3.4) and Hologic QDR-1000 dual-energy X-ray absorptiometers have comparable normal reference databases for the spine and femur of white UK and USA subjects. After conversion for systematic differences in absolute bone density values between the two systems, the reference databases were very similar for the spine in young subjects, but there were clear differences in the femur databases of young females and males of all ages. These differences were confirmed by comparing the percent age-matched and young values determined by the two systems for subjects scanned on both systems. Thus the diagnosis and management of a patient could differ, depending on the system used for the bone density measurements.
9
Performance evaluation of a dual-energy X-ray bone densitometer
Richard B. Mazess, B. Collick, J. Trempe et al. · Calcified Tissue International · 1989 · 353 citations
Sex differences in peak adult bone mineral density
Paul J. Kelly, Lance Twomey, Philip N. Sambrook et al. · Journal of Bone and Mineral Research · 1990 · 151 citations