IEEE Electron Device Letters · 2011 · 22 citations · 8 references
EngineeringBioengineeringBiomedical EngineeringBiomemsBiosensorsSensorsBiological SensorsBiomedical SensorsBiosensor DesignChemical SensorsMicrobial BiosensorsNanosensorsBacterial BiosensorsBacterial SensorsBiosensor CharacterizationBiosensor DevelopmentElectrical PropertiesMicrocantilever BiosensorsNanobiosensorsStimulant Detection
In this work, MOSFET-embedded cantilevers are configured as microbial sensors for detection of anthrax simulants, Bacillus thuringiensis. Anthrax simulants attached to the chemically treated gold-coated cantilever cause changes in the MOSFET drain current due to the bending of the cantilever which indicates the detection of anthrax simulant. Electrical properties of the anthrax simulant are also responsible for the change in the drain current. The test results suggest a detection range of 10 μL of stimulant test solution (a suspension population of 1.3 × 10 <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">7</sup> colony-forming units/mL diluted in 40% ethanol and 60% deionized water) with a linear response of 31 μA/μL.
8
Christine Faille, Grégoire Tauveron, C. Le Gentil‐Lelièvre et al. · Journal of Food Protection · 2007 · 73 citations · Full text