The Journal of Rehabilitation Research and Development · 2004 · 132 citations · 28 references
Life SciencesNeurosciencePhysiologyElectrophysiologyOtologyVestibular DisordersVestibular Evoked Myogenic PotentialsVestibular FunctionVestibular NeuroscienceVestibular PhysiologyVestibular SystemVestibular TestingElectromyography Signal AcquisitionRecordingsSaccular FunctionVestibular Neurophysiology
Vestibular-evoked myogenic potentials (VEMPs) are proposed as a reliable test to supplement the current vestibular test battery by providing diagnostic information about saccular and/or inferior vestibular nerve function. VEMPs are short-latency electromyograms (EMGs) evoked by high-level acoustic stimuli and recorded from surface electrodes over the tonically contracted sternocleidomastoid muscle. VEMP amplitude is influenced by the EMG level, which must be controlled. This study examined the ability of subjects to achieve the EMG target levels over a range of target levels typically used during VEMP recordings. In addition, the influence of target EMG level on the latency and amplitude of the click- and tone-evoked VEMP was examined. The VEMP amplitude increased as a function of EMG target level, and the latency remained constant. EMG target levels ranging from 30 microV to 50 microV are suggested for clinical application of the VEMP.
28
Toshihisa Murofushi, G. Michael Halmagyi, R.A. Yavor et al. · Archives of Otolaryngology - Head and Neck Surgery · 1996 · 369 citations
Medicine, Neurology, Otolaryngology +12
Responses of guinea pig primary vestibular neurons to clicks
Toshihisa Murofushi, Ian S. Curthoys, AnnN. Topple et al. · Experimental Brain Research · 1995 · 349 citations