American Journal of Neuroradiology · 2011 · 53 citations · 28 references
MedicineNeurologyNeuroscienceBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyBrain ImagingCerebrovascular ImagingEpilepsy ImagingMagnetic Resonance ImagingFalse-positive ResultsInversion Recovery SequencesMagnetic Resonance CisternographyNeuroimaging Of Subarachnoid HemorrhagePulse Inversion ImagingSensitivityThree-dimensional Cranial Imaging
BACKGROUND AND PURPOSE: FLAIR images are highly sensitive for SAH. However, CSF flow artifacts caused by conventional FLAIR can produce false-positive results. Here, we compare 3D and 3D FLAIR sequences, focusing on their potential for containing these artifacts and their sensitivity and specificity for detection of SAHs. MATERIALS AND METHODS: We evaluated the following 4 FLAIR sequences: 1) 2D FLAIR at 1.5T, 2) 2D FLAIR, 3) 2D PROPELLER-FLAIR, and 4) 3D Cube-FLAIR at 3T. All sequences were performed in 5 healthy volunteers; sequences 2 and 4 were also performed under routine conditions in 10 patients with focal epilepsy and in 10 patients with SAH. Two neuroradiologists independently conducted the analysis. The presence of flow artifacts in the ventricles and cisterns of healthy volunteers and patients with epilepsy was evaluated and scored on a 4-point scale. Mean values were calculated and compared by using paired t tests. Sensitivity and specificity for SAH detection in sequences 2 and 4 were determined. RESULTS: Cube-FLAIR showed almost no CSF artifacts in the volunteers and the patients with epilepsy; therefore, it was superior to any other FLAIR (P < .001). Sensitivity and specificity of SAH detection by 3T FLAIR were 58.3% and 89.4%, respectively, whereas Cube-FLAIR had a sensitivity of 95% and a specificity of 100%. CONCLUSIONS: Cube-FLAIR allows FLAIR imaging with almost no CSF artifacts and is, thus, particularly useful for SAH detection.
28
The Measurement of Observer Agreement for Categorical Data
J. Richard Landis, Gary G. Koch · Biometrics · 1977 · 78.9K citations · Full text
Comparison of MRI and CT for Detection of Acute Intracerebral Hemorrhage
Chelsea S. Kidwell · JAMA · 2004 · 818 citations
Reed F. Busse, Hari Hariharan, Anthony T. Vu et al. · Magnetic Resonance in Medicine · 2006 · 329 citations · Full text
Acute subarachnoid hemorrhage: MR imaging with fluid-attenuated inversion recovery pulse sequences.
Kyo Noguchi, Takahiro Ogawa, A Inugami et al. · Radiology · 1995 · 257 citations