Magnetic Resonance in Medicine · 1999 · 917 citations · 15 references
MedicineBioengineeringRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyCardiac ImagingMagnetic Resonance Image ReconstructionMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMedical ImagingMotion CorrectionCardiac Motion ImagingHead Motion CorrectionMotion ArtifactsMotion-corrected Image ReconstructionMotion-corrected ReconstructionPatient Motion CorrectionRespiratory Motion CompensationRespiratory Motion Correction
The study presents a motion‑correction method that integrates data acquisition and reconstruction. The method uses PROPELLER MRI, acquiring concentric rectangular strips rotated around the k‑space origin, with repeated central sampling to correct spatial inconsistencies, reject through‑plane motion, and reduce artifacts through low‑frequency averaging. The technique successfully corrected bulk head motion and respiratory motion in ungated cardiac imaging. Published in Magn Reson Med 42:963‑969 (1999).
A method for motion correction, involving both data collection and reconstruction, is presented. The PROPELLER MRI method collects data in concentric rectangular strips rotated about the k-space origin. The central region of k-space is sampled for every strip, which (a) allows one to correct spatial inconsistencies in position, rotation, and phase between strips, (b) allows one to reject data based on a correlation measure indicating through-plane motion, and (c) further decreases motion artifacts through an averaging effect for low spatial frequencies. Results are shown in which PROPELLER MRI is used to correct for bulk motion in head images and respiratory motion in nongated cardiac images. Magn Reson Med 42:963-969, 1999.
15
Fast Spiral Coronary Artery Imaging
Craig H. Meyer, Bob S. Hu, Dwight G. Nishimura et al. · Magnetic Resonance in Medicine · 1992 · 885 citations
Medicine, Cardiology, Radiology +14