Enhanced Sensitivity for Selected Reaction Monitoring Mass Spectrometry-based Targeted Proteomics Using a Dual Stage Electrodynamic Ion Funnel Interface

Mahmud Hossain, David T. Kaleta, Errol Robinson, Tao Liu, Rui Zhao, Jason S. Page, Ryan Kelly, Ronald Moore, Keqi Tang, David Camp,

Molecular & Cellular Proteomics · 2010 · 52 citations · 24 references

DOIFull text

Open access

Concepts

Abstract

Selected reaction monitoring mass spectrometry (SRM-MS) is playing an increasing role in quantitative proteomics and biomarker discovery studies as a method for high throughput candidate quantification and verification. Although SRM-MS offers advantages in sensitivity and quantification compared with other MS-based techniques, current SRM technologies are still challenged by detection and quantification of low abundance proteins (e.g. present at ∼10 ng/ml or lower levels in blood plasma). Here we report enhanced detection sensitivity and reproducibility for SRM-based targeted proteomics by coupling a nanospray ionization multicapillary inlet/dual electrodynamic ion funnel interface to a commercial triple quadrupole mass spectrometer. Because of the increased efficiency in ion transmission, significant enhancements in overall signal intensities and improved limits of detection were observed with the new interface compared with the original interface for SRM measurements of tryptic peptides from proteins spiked into non-depleted mouse plasma over a range of concentrations. Overall, average SRM peak intensities were increased by ∼70-fold. The average level of detection for peptides also improved by ∼10-fold with notably improved reproducibility of peptide measurements as indicated by the reduced coefficients of variance. The ability to detect proteins ranging from 40 to 80 ng/ml within mouse plasma was demonstrated for all spiked proteins without the application of front-end immunoaffinity depletion and fractionation. This significant improvement in detection sensitivity for low abundance proteins in complex matrices is expected to enhance a broad range of SRM-MS applications including targeted protein and metabolite validation. Selected reaction monitoring mass spectrometry (SRM-MS) is playing an increasing role in quantitative proteomics and biomarker discovery studies as a method for high throughput candidate quantification and verification. Although SRM-MS offers advantages in sensitivity and quantification compared with other MS-based techniques, current SRM technologies are still challenged by detection and quantification of low abundance proteins (e.g. present at ∼10 ng/ml or lower levels in blood plasma). Here we report enhanced detection sensitivity and reproducibility for SRM-based targeted proteomics by coupling a nanospray ionization multicapillary inlet/dual electrodynamic ion funnel interface to a commercial triple quadrupole mass spectrometer. Because of the increased efficiency in ion transmission, significant enhancements in overall signal intensities and improved limits of detection were observed with the new interface compared with the original interface for SRM measurements of tryptic peptides from proteins spiked into non-depleted mouse plasma over a range of concentrations. Overall, average SRM peak intensities were increased by ∼70-fold. The average level of detection for peptides also improved by ∼10-fold with notably improved reproducibility of peptide measurements as indicated by the reduced coefficients of variance. The ability to detect proteins ranging from 40 to 80 ng/ml within mouse plasma was demonstrated for all spiked proteins without the application of front-end immunoaffinity depletion and fractionation. This significant improvement in detection sensitivity for low abundance proteins in complex matrices is expected to enhance a broad range of SRM-MS applications including targeted protein and metabolite validation. Although mass spectrometry (MS)-based proteomics is a promising high throughput technology for biomarker discovery and validation (1.Rifai N. Gillette M.A. Carr S.A. Protein biomarker discovery and validation: the long and uncertain path to clinical utility.Nat. Biotechnol. 2006; 24: 971-983Crossref PubMed Scopus (1383) Google Scholar, 2.Lu M. Faull K.F. Whitelegge J.P. He J. Shen D. Saxton R.E. Chang H.R. Proteomics and mass spectrometry for cancer biomarker discovery.Biomarker Insights. 2007; 2: 347-360Crossref PubMed Google Scholar, 3.Schiess R. Wollscheid B. Aebersold R. Targeted proteomic strategy for clinical biomarker discovery.Mol. Oncol. 2009; 3: 33-44Crossref PubMed Scopus (295) Google Scholar, 4.Hanash S.M. Pitteri S.J. Faca V.M. Mining the plasma proteome for cancer biomarkers.Nature. 2008; 452: 571-579Crossref PubMed Scopus (691) Google Scholar, 5.Qian W.J. Jacobs J.M. Liu T. Camp 2nd, D.G. Smith R.D. Advances and challenges in liquid chromatography-mass spectrometry-based proteomics profiling for clinical applications.Mol. Cell. Proteomics. 2006; 5: 1727-1744Abstract Full Text Full Text PDF PubMed Scopus (310) Google Scholar), only a handful of cancer biomarkers have been approved by the United States Food and Drug Administration for clinical use in the last decade (6.Ludwig J.A. Weinstein J.N. Biomarkers in cancer staging, prognosis and treatment selection.Nat. Rev. Cancer. 2005; 5: 845-856Crossref PubMed Scopus (1346) Google Scholar, 7.Anderson N.L. Anderson N.G. The human plasma proteome: history, character, and diagnostic prospects.Mol. Cell. Proteomics. 2002; 1: 845-867Abstract Full Text Full Text PDF PubMed Scopus (3586) Google Scholar). Assuming that low abundance biomarkers do exist in the biofluids to be studied, the success of biomarker discovery efforts primarily depends on the sensitivity, accuracy, and robustness of the measurement technologies; the quality and size of patient cohorts and clinical samples and execution within the context of an overall difficult and expensive path to clinical application that encompasses discovery, verification, and validation stages (1.Rifai N. Gillette M.A. Carr S.A. Protein biomarker discovery and validation: the long and uncertain path to clinical utility.Nat. Biotechnol. 2006; 24: 971-983Crossref PubMed Scopus (1383) Google Scholar, 5.Qian W.J. Jacobs J.M. Liu T. Camp 2nd, D.G. Smith R.D. Advances and challenges in liquid chromatography-mass spectrometry-based proteomics profiling for clinical applications.Mol. Cell. Proteomics. 2006; 5: 1727-1744Abstract Full Text Full Text PDF PubMed Scopus (310) Google Scholar, 8.Polanski M. Anderson N.L. A list of candidate cancer biomarkers for targeted proteomics.Biomarker Insights. 2007; 1: 1-48PubMed Google Scholar, 9.Ye X. Blonder J. Veenstra T.D. Targeted proteomics for validation of biomarkers in clinical samples.Brief. Funct. Genomic. Proteomic. 2009; 8: 126-135Crossref PubMed Scopus (34) Google Scholar, 10.Carr S.A. Anderson L. Protein quantitation through targeted mass spectrometry: the way out of biomarker purgatory?.Clin. Chem. 2008; 54: 1749-1752Crossref PubMed Scopus (124) Google Scholar). A multiplexed assay platform increasingly considered for biomarker verification is selected reaction monitoring (SRM) 1The abbreviations used are:SRMselected reaction monitoringLODlimit of detectionCVcoefficient of varianceQqQtriple quadrupoleRFradio frequencyDCdirect currentQCquality control. by tandem mass spectrometry using e.g. a triple quadrupole (QqQ) mass spectrometer to attain high throughput quantitative measurements of targeted proteins in complex matrices (1.Rifai N. Gillette M.A. Carr S.A. Protein biomarker discovery and validation: the long and uncertain path to clinical utility.Nat. Biotechnol. 2006; 24: 971-983Crossref PubMed Scopus (1383) Google Scholar, 11.Anderson L. Hunter C.L. Quantitative mass spectrometric multiple reaction monitoring assays for major plasma proteins.Mol. Cell. Proteomics. 2006; 5: 573-588Abstract Full Text Full Text PDF PubMed Scopus (1085) Google Scholar, 12.Keshishian H. Addona T. Burgess M. Kuhn E. Carr S.A. Quantitative, multiplexed assays for low abundance proteins in plasma by targeted mass spectrometry and stable isotope dilution.Mol. Cell. Proteomics. 2007; 6: 2212-2229Abstract Full Text Full Text PDF PubMed Scopus (575) Google Scholar). selected reaction monitoring limit of detection coefficient of variance triple quadrupole radio frequency direct current quality control. SRM utilizes two stages of mass filtering by selecting a specific analyte ion of interest (precursor ion) in the first stage followed by a specific fragment ion derived from the precursor (fragment ion) filter in the second stage after collision-activated dissociation. Typically, several transitions (precursor/fragment ion pairs) are monitored for greater selectivity and confidence in a targeted peptide assay, and large numbers of peptides can be monitored during a single LC-MS/MS analysis. The two-stage mass selection by individual quadrupoles enables more rapid and continuous monitoring of specific ions derived from analytes of interest such as peptides and leads to significantly enhanced detection sensitivity and quantitative accuracy compared with broad (i.e. non-targeted) LC-MS or LC-MS/MS measurements (11.Anderson L. Hunter C.L. Quantitative mass spectrometric multiple reaction monitoring assays for major plasma proteins.Mol. Cell. Proteomics. 2006; 5: 573-588Abstract Full Text Full Text PDF PubMed Scopus (1085) Google Scholar, 12.Keshishian H. Addona T. Burgess M. Kuhn E. Carr S.A. Quantitative, multiplexed assays for low abundance proteins in plasma by targeted mass spectrometry and stable isotope dilution.Mol. Cell. Proteomics. 2007; 6: 2212-2229Abstract Full Text Full Text PDF PubMed Scopus (575) Google Scholar). Both the sensitivity and selectivity of SRM-MS make this technique well suited for the targeted detection and quantification of low abundance proteins in highly complex biofluids (13.Lange V. Picotti P. Domon B. Aebersold R. Selected reaction monitoring for quantitative proteomics: a tutorial.Mol. Syst. Biol. 2008; 4: 222Crossref PubMed Scopus (1142) Google Scholar, 14.Kuzyk M.A. Smith D. Yang J. Cross T.J. Jackson A.M. Hardie D.B. Anderson N.L. Borchers C.H. Multiple reaction monitoring-based, multiplexed, absolute quantitation of 45 proteins in human plasma.Mol. Cell. Proteomics. 2009; 8: 1860-1877Abstract Full Text Full Text PDF PubMed Scopus (450) Google Scholar, 15.Kuhn E. Addona T. Keshishian H. Burgess M. Mani D.R. Lee R.T. Sabatine M.S. Gerszten R.E. Carr S.A. Developing multiplexed assays for troponin I and interleukin-33 in plasma by peptide immunoaffinity enrichment and targeted mass spectrometry.Clin. Chem. 2009; 55: 1108-1117Crossref PubMed Scopus (211) Google Scholar, 16.DeSouza L.V. Romaschin A.D. Colgan T.J. Siu K.W. Absolute quantification of potential cancer markers in clinical tissue homogenates using multiple reaction monitoring on a hybrid triple quadrupole/linear ion trap tandem mass spectrometer.Anal. Chem. 2009; 81: 3462-3470Crossref PubMed Scopus (71) Google Scholar). The precision and reproducibility of SRM-based measurements of proteins in plasma across different laboratories have recently been assessed (17.Addona T.A. Abbatiello S.E. Schilling B. Skates S.J. Mani D.R. Bunk D.M. Spiegelman C.H. Zimmerman L.J. Ham A.J. Keshishian H. Hall S.C. Allen S. Blackman R.K. Borchers C.H. Buck C. Cardasis H.L. Cusack M.P. Dodder N.G. Gibson B.W. Held J.M. Hiltke T. Jackson A. Johansen E.B. Kinsinger C.R. Li J. Mesri M. Neubert T.A. Niles R.K. Pulsipher T.C. Ransohoff D. Rodriguez H. Rudnick P.A. Smith D. Tabb D.L. Tegeler T.J. Variyath A.M. Vega-Montoto L.J. Wahlander A. Waldemarson S. Wang M. Whiteaker J.R. Zhao L. Anderson N.L. Fisher S.J. Liebler D.C. Paulovich A.G. Regnier F.E. Tempst P. Carr S.A. Multi-site assessment of the precision and reproducibility of multiple reaction monitoring-based measurements of proteins in plasma.Nat. Biotechnol. 2009; 27: 633-641Crossref PubMed Scopus (874) Google Scholar). Despite its promise, present SRM measurements still do not provide sufficient sensitivity for reliable detection and quantification of low abundance proteins in biofluids (e.g. present in plasma at ∼10 ng/ml or lower levels) primarily because of factors related to high sample complexity and the large dynamic range of relative protein abundances (7.Anderson N.L. Anderson N.G. The human plasma proteome: history, character, and diagnostic prospects.Mol. Cell. Proteomics. 2002; 1: 845-867Abstract Full Text Full Text PDF PubMed Scopus (3586) Google Scholar, 18.Xu X. Veenstra T.D. Analysis of biofluids for biomarker research.Proteomics Clin. Appl. 2008; 2: 1403-1412Crossref PubMed Scopus (19) Google Scholar, 19.Qian W.J. Kaleta D.T. Petritis B.O. Jiang H. Liu T. Zhang X. Mottaz H.M. Varnum S.M. Camp 2nd, D.G. Huang L. Fang X. Zhang W.W. Smith R.D. Enhanced detection of low abundance human plasma proteins using a tandem IgY12-SuperMix immunoaffinity separation strategy.Mol. Cell. Proteomics. 2008; 7: 1963-1973Abstract Full Text Full Text PDF PubMed Scopus (169) Google Scholar). Given sufficient selectivity, the sensitivity achievable is generally related to the peptide MS and MS/MS signal intensities obtained. One of the key factors limiting peptide MS intensities is the significant ion losses encountered between the electrospray ionization (ESI) source and the interface to the mass spectrometer. In typical LC-ESI-MS interfaces, the mass spectrometer inlet (e.g. heated capillary followed by a skimmer) presently provides total ion utilization and ion transmission efficiencies on the order of ∼1% (20.Page J.S. Kelly R.T. Tang K. Smith R.D. Ionization and transmission efficiency in an electrospray ionization-mass spectrometry interface.J. Am. Soc. Mass 2007; PubMed Scopus Google to a of ion from the ion source into the inlet and transmission of ions the first reduced stage of the mass spectrometer. The electrodynamic ion funnel R.T. J.S. Tang K. Smith R.D. The ion and Rev. Google Scholar), been to and ions to the high of the mass is expected to provide a large to SRM The original ion funnel interfaces, at a of were to enhance signal intensities for a of MS S.A. Tang K. Anderson D.C. H.R. Smith R.D. A ion funnel for ions at using electrospray ionization mass Mass Scopus Google Scholar, S.A. A. D.C. Anderson H.R. Smith R.D. of an improved electrodynamic ion funnel interface for electrospray ionization mass Chem. PubMed Scopus Google Scholar, T. R. D.C. Anderson H.R. Smith R.D. and of a new electrodynamic ion Chem. PubMed Scopus Google by the Although ion transmission to high losses at the interface a high ion funnel interface of at a of was Tang K. Smith R.D. mass spectrometer sensitivity using a electrodynamic ion funnel interface.J. Am. Soc. Mass 2006; PubMed Scopus Google Scholar). The greater and more ion from through a multicapillary a ion funnel interface a high ion funnel with a heated multicapillary inlet followed by a ion funnel at highly ion from to high is In this we report the enhanced sensitivity and reproducibility of SRM-based targeted proteomics measurements by a stage electrodynamic ion funnel interface that a multicapillary inlet with a triple quadrupole mass spectrometer. A of measurements were using mouse plasma samples spiked with of tryptic peptides from proteins to the in detection sensitivity and reproducibility by this interface relative to a capillary A ∼10-fold improvement in the limit of detection as well as improved measurement reproducibility was The proteins and and the peptides fragment and were from and were from was from and the mouse plasma was from The mouse plasma protein was as by protein assay were for of the E. and the proteins were using the as that used to the mouse of proteins was and reduced in for at Protein were with 40 for at The were with and was at a of sample was at for a and with of were from the with of and in of The peptide in of the was by protein assay and of the were to a new with of of the for SRM were by different of the of protein into a mouse plasma tryptic to The were in to the protein in the original mouse mouse samples were to a of total The samples were into in liquid and at protein levels used are in ng/ml protein in original plasma sample total and mass to mass level relative to total plasma to mass in a new samples were using a high capillary Tang K. M.A. Zhao R. Shen Kelly R.T. H.R. Smith R.D. capillary LC-MS for throughput in proteomic Chem. 2008; PubMed Scopus Google on to a triple quadrupole mass spectrometer Fisher were using the capillary was by into a capillary This capillary was a to a R.T. J.S. Tang K. Smith R.D. and for ionization mass Chem. 2006; PubMed Scopus Google in The of in and in of the peptide the an was by increasing the from to over The inlet capillary of the mass spectrometer was at with an of that was to the were using the interface or the ion funnel In the ion funnel Tang K. Smith R.D. mass spectrometer sensitivity using a electrodynamic ion funnel interface.J. Am. Soc. Mass 2006; PubMed Scopus Google Scholar), high funnel and low funnel were used in tandem Both ion of was and by The of of the first was and the of the last 45 from to The total of the high ion funnel was in radio frequency for ion transmission in the of the high ion A was to to an potential well that the a the ions through the ion The high ion funnel an of and a of on the last ion funnel The low ion funnel of and on the first and last ion funnel The two ion were by a The also a with a of from the ion funnel the funnel T. Tang K. H.R. Smith R.D. A multicapillary inlet electrodynamic ion funnel interface for improved sensitivity using ion Chem. PubMed Scopus Google Scholar). The potential of the and limit were at and for the high ion funnel and and for the low ion The ion funnel of a multicapillary significantly increased ion transmission into the mass spectrometer relative to the single capillary inlet The multicapillary inlet was using an of and into a of The of the was also in the ion funnel using the two to the of the original was used in with a second a and with a to the in the ion funnel The was to in the low ion funnel and SRM transitions for protein were and were by direct of a of protein in a of A and at an of transitions were targeted for and the two peptides were selected for of the two transitions for quality peptide were also The selection of peptides and transitions was on the in the mass SRM a of and a of were Both and were to a peak of at were in ion were using with detection and were using the with on the spiked samples 40 and using the of peak and a of In were to a peak of and of for the more of the were and the from the peptide peak was fragment and were used as peptides for markers and for monitoring the overall of the The was by the sample in this a of at for the was not used in the peak The peak for peptide were by the using the two spiked and and the observed relative to the as a peak for the lower concentrations. peak at the lower was on The of from the was within of the average observed from the two concentrations. peak were (i.e. within to the average of peak from the two concentrations. peak with analyte This the in SRM-MS measurements achievable the increased signal levels using a more the of the ion funnel interface to the mass spectrometer. The high ion funnel used a inlet of the single capillary inlet for a low ion funnel The greater efficiency with the inlet significantly more ions to be into the mass spectrometer from the overall ion transmission efficiency to the mass spectrometer. the ion funnel interface for SRM we selected proteins that were spiked into a of mouse plasma at Protein from to ng/ml in mouse plasma as in I A between the ion funnel interface and the interface was by of samples using interface peptides spiked into sample at a level of to as of the two peptides were selected from transitions for selected peptide were the average in SRM peak with the multicapillary inlet/dual ion funnel interface relative to observed with the The for the peptides in this from of peak were enhanced with the ion funnel interface with peak ranging from to for individual from the plasma was observed with of the its peak for all of the SRM transitions are in Overall, the observed in SRM peak intensities are in with on for MS signal were observed a low ion funnel interface was to a triple quadrupole mass spectrometer S.A. Tang K. Anderson D.C. H.R. Smith R.D. A ion funnel for ions at using electrospray ionization mass Mass Scopus Google Scholar), and an in signal relative to a low ion funnel was using a ion funnel with a multicapillary inlet in a mass spectrometer Tang K. Smith R.D. mass spectrometer sensitivity using a electrodynamic ion funnel interface.J. Am. Soc. Mass 2006; PubMed Scopus Google Scholar). overall of was observed a ion funnel was compared with an R.T. J.S. Zhao R. W.J. Mottaz H.M. Tang K. Smith R.D. in ion transmission efficiency and with capillary Chem. 2008; PubMed Scopus Google Scholar). was to the increased signal intensities enhanced detection of the low abundance spiked proteins in mouse an improved lower for targeted the SRM peak observed for peptides at different analyte and all transitions with the and ion funnel of SRM for the high for peptide and for peptide of at different levels ng/ml and are in the enhanced peak intensities in peak and improved to lower the and coefficient of variance for the transitions for peptide with the two interface The was as the that was observed with signal The that sensitivity and detection limit are improved by using the ion funnel the ion funnel of peptides were at ng/ml with the the for the peptides were ng/ml with the of peptide at 80 the overall improvement in for all targeted peptides using the ion funnel interface was ∼10-fold with only for peptide because of between the peptide of interest and a peak that quantification of proteins present at The for all transitions in are as of observed ion using ion funnel and for two transitions of peptide and of peptide from protein peak ion of peptide for transitions peptide ion ion ion funnel at for selected transitions at the of the for the interface with the ion funnel from the were observed to limit the detection and quantification of the of from the were observed to limit the detection and quantification of the of from the were observed to limit the detection and quantification of the of from the were observed to limit the detection and quantification of the of from the were observed to limit the detection and quantification of the of for selected transitions at the of the for the interface with the ion funnel from the were observed to limit the detection and quantification of the of in a new In to enhanced sensitivity, we also observed improved Both were first compared for the reproducibility of transitions of peptide using the Here interface levels were used for reproducibility was also for only the ion funnel interface at its that the for the selected transitions with the ion funnel interface are lower with the interface for the of samples at specific the the of the selected transitions at its lower detection level from to as high as with an average of the with the ion funnel interface range from to with an average of for peptides at the as the The for of peptides are ranging from to with an average of for peptides at the funnel interface lower detection Although the in sensitivity and reproducibility are because of the increased ion transmission through the new such significant observed are to the selectivity by the two-stage mass filtering in SRM In typical LC-MS the detection of analyte is by is in Because the ion funnel interface overall ion transmission without the in LC-MS measurements do not have a on LC-MS analyte detection sensitivity as observed Tang K. Smith R.D. mass spectrometer sensitivity using a electrodynamic ion funnel interface.J. Am. Soc. Mass 2006; PubMed Scopus Google Scholar). the two-stage mass filtering for the ions at different quadrupoles in SRM-MS measurements significantly the to a low level by only analyte signal to a the detection of a analyte by SRM is by are sufficient analyte to the of the SRM-MS an platform for of the increased ion transmission of the ion funnel In the SRM measurement the increased ion transmission in enhanced signal intensities without a increased level to the mass to a lower limit of detection observed the increased improved measurement reproducibility because the increased signal intensities the of is the of overall at low signal measurement sensitivity, in measurements is the of with the analytes of interest as observed in make difficult to analyte peak for Although the use of stable peptides is to analyte on in separation and that the sample complexity be to enhance the selectivity of overall In this we demonstrated that the of a new interface that multicapillary and electrodynamic ion on a triple quadrupole mass spectrometer significantly SRM sensitivity and reproducibility as assessed by the and of peptides spiked into non-depleted mouse improvement in ion transmission efficiency to the mass spectrometer with the new interface was by the of a average in SRM peak intensities for all peptides relative to the The increased ion signal intensities in a lower analyte and The improved on average by ∼10-fold for the selected all selected reproducibility was by coefficients of variance from ng/ml proteins spiked into mouse plasma were without the application of front-end immunoaffinity depletion and using this new that the coupling of this ion funnel interface with a triple quadrupole mass spectrometer broad applications targeted measurements of low abundance analytes (e.g. proteins or in complex samples because of the enhanced sensitivity and coupling with front-end single R. C.L. Huang Anderson N.L. S. immunoaffinity an a of the human plasma 3: PubMed Scopus Google or stage W.J. Kaleta D.T. Petritis B.O. Jiang H. Liu T. Zhang X. Mottaz H.M. Varnum S.M. Camp 2nd, D.G. Huang L. Fang X. Zhang W.W. Smith R.D. Enhanced detection of low abundance human plasma proteins using a tandem IgY12-SuperMix immunoaffinity separation strategy.Mol. Cell. Proteomics. 2008; 7: 1963-1973Abstract Full Text Full Text PDF PubMed Scopus (169) Google immunoaffinity depletion and SRM-MS with this new interface is expected to measurements at levels for proteins in human

References

24