The presence of Pseudomonas aeruginosa in the dental unit waterline systems of teaching clinics

AS Al‐Hiyasat, Showgy Y. Ma’ayeh

International Journal of Dental Hygiene · 2007 · 54 citations · 37 references

Concepts

Abstract

OBJECTIVE: The objective of this study was to evaluate the extent of Pseudomonas aeruginosa contamination of Dental Unit Water (DUW) at a Dental Teaching Center in Jordan. METHODS: Water samples were collected from 30 dental units, 10 from each of three teaching clinics, namely conservative dentistry, periodontology, and prosthodontics. Samples were collected from the outlet of the air/water syringe, high-speed handpiece and water cup filler, at the beginning of the working day (before use), after 2 min flushing, and at midday. RESULTS: P. aeruginosa was detected in 86.7% (26/30) of the dental units at the beginning of the working day, and in 73.3% (22/30) after 2 min of flushing and at midday. Conservative dentistry units had the highest counts, followed by periodontology and prosthodontics (P<0.05). Overall, the highest counts (log10 count CFU ml-1) were at the beginning of the working day (1.38+/-1.05), and the lowest counts after flushing for 2 min (1.10+/-1.03), and higher numbers were seen again at midday (1.15+/-1.04) (P<0.05). CONCLUSIONS: 86.7% of the dental units were contaminated with P. aeruginosa, the conservative dentistry units had the highest amount of contamination. Flushing the DUW for 2 min significantly reduced the counts of P. aeruginosa.