Applied Optics · 1978 · 21 citations · 13 references
EngineeringElectrical EngineeringRadiation MeasurementSemiconductor DevicesSensorsCalibrationDevice CharacterizationEquipment CalibrationRadiometric CalibrationTemperature MeasurementTemperature SensorsAbsorber MaterialsPrecisionRadiation Instrument CalibrationRadiometer CalibrationRadiometer DesignSystematic UncertaintiesThermal CalibrationTime Constant
The fabrication and principal characteristics of an electrically calibrated absolute radiometer are described. The receiver substrate incorporates an evaporated chromium-nickel thermopile, an electrical shield, a copper thermal diffuser disk, an evaporated chromium electrical heating element, and a goldblack absorber. All insulating layers are evaporated silicon monoxide; the thermopile and heater are made by a photoetching process. The performances of several radiometers are discussed. For example, a 28-junction version has a responsivity and NEP of 93 mV/W and 50 nW, respectively, in air, with a time constant of 15 sec and a surface responsivity uniform to better than 1%. The radiometers require only a few corrections of small magnitude. An analysis of sources of error and residual uncertainties shows that the over-all precision of this type of radiometer is at least 0.5% for a power level of 50 microW. Comparative radiometric measurements are described that support this claim.
13
Spectral irradiance standard for the ultraviolet: the deuterium lamp
R. D. Saunders, W. R. Ott, J. M. Bridges · Applied Optics · 1978 · 57 citations
Natural Sciences, Physics, Optics +14
Diffraction Losses in Radiometry and Photometry
W. R. Blevin · Metrologia · 1970 · 52 citations