Publication | Closed Access
An improved amperometric immunosensor for the detection and enumeration of protein A-bearing Staphylococcus aureus
28
Citations
9
References
1990
Year
An amperometric immunosensor specific to the protein A of Staphylococcus aureus, was developed using the direct electrochemical detection of phenol produced by alkaline phosphatase from phenylphosphate. The immunosensor could detect protein A at 0.01 ng/ml and could reliably detect and quantify pure cultures of protein A-bearing Staph. aureus above 10(3) cfu/ml. A similar sensitivity of detection was obtained with samples of beef and milk.