Rapid Communications in Mass Spectrometry · 1992 · 21 citations · 27 references
Natural SciencesChemistryPharmaceutical AnalysisAnalytical ChemistryMass Spectrometric AnalysisSpectroscopyAtmospheric IonizationInstrumental AnalysisIonization ProcessesIonization TechniquesMass AnalyzersMass SpectrometryAnalytical Liquid ChromatographyAtmospheric PressureIon Trap Mass SpectrometryIonizationIonization Interface DesignPotential DifferenceQuadrupole Mass SpectrometryQuaternary Ammonium Compounds
Abstract A novel interface is described that effects the sampling and mass analysis of ions formed at atmospheric pressure into a Saturn II bench‐top quadrupole ion‐trap mass spectrometer. The standard high‐vacuum pumping system of the system is unchanged, bit a 960L/min rotary pump is incorporated to pump the first vacuum stage of the atmospheric‐pressure ion sampling interface. An ‘Electric skimmer’ ion focusing region in the first vacuum turbomolecular pump of the Sasturn II maintains high vaccuum at 4×10 −4 Torr in the analyzer region while helium damping gas is introduced. By vaarying the potential difference between selected lenses in the first vacuum region the ‘declustering’ energy may be varied to decluster effectively ion/molecule dissociation mass spectra. The on‐line analysis of quaternary ammonium drugs by liquid chromatography of 5 ng and 150 pg, respectively.
27
Electrospray ionization–principles and practice
John B. Fenn, Matthias Mann, Chin Kai Meng et al. · Mass Spectrometry Reviews · 1990 · 1.5K citations