IEEE Transactions on Biomedical Engineering · 2009 · 30 citations · 25 references
In this paper, novel procedures were developed using a thin (0.17 mm) tactile piezoresistive pressure sensor mounted on a rigid contact lens to measure upper eyelid pressure. A hydrostatic calibration system was constructed, and the influence of conditioning (prestressing), drift (continued increasing response with a static load), and temperature variations on the response of the sensor were examined. To optimally position the sensor-contact lens combination under the upper eyelid margin, an in vivo measurement apparatus was constructed. Calibration gave a linear relationship between raw sensor output and actual pressure units for loads between 1 and 10 mmHg ( R(2) = 0.96 ). Conditioning the sensor prior to use regulated the measurement response, and sensor output stabilized about 10 s after loading. While sensor output drifts slightly over several hours, it was not significant beyond the measurement time of 1 min used for eyelid pressure. The error associated with calibrating at room temperature but measuring at ocular surface temperature led to a very small overestimation of pressure. Eyelid pressure readings were observed when the upper eyelid was placed on the sensor, and removed during a recording. When the eyelid pressure was increased by pulling the lids tighter against the eye, the readings from the sensor significantly increased.