Publication | Closed Access
Respiratory Motion Correction in 3-D PET Data With Advanced Optical Flow Algorithms
170
Citations
35
References
2008
Year
Computed TomographyOptical Flow AlgorithmsImage ReconstructionEngineeringPet-mriAdvanced ImagingBiomedical EngineeringPositron Emission TomographyImage Analysis3-D Pet DataCt ScanPhoton-counting Computed TomographyRespiratory Motion CorrectionNuclear MedicineRadiologyHealth SciencesMedical ImagingPet ImagesInverse ProblemsMedical Image ComputingComputer VisionBiomedical Imaging3D Imaging
The problem of motion is well known in positron emission tomography (PET) studies. The PET images are formed over an elongated period of time. As the patients cannot hold breath during the PET acquisition, spatial blurring and motion artifacts are the natural result. These may lead to wrong quantification of the radioactive uptake. We present a solution to this problem by respiratory-gating the PET data and correcting the PET images for motion with optical flow algorithms. The algorithm is based on the combined local and global optical flow algorithm with modifications to allow for discontinuity preservation across organ boundaries and for application to 3-D volume sets. The superiority of the algorithm over previous work is demonstrated on software phantom and real patient data.