Journal of Clinical Microbiology · 2000 · 16 citations · 23 references
Public HealthInfectious DiseasesMolecular BiologyMolecular DiagnosticsMolecular VirologyVirologyAntiretroviral TherapyHiv DiagnosisHiv TestingInfectious Disease Molecular TestingRna-based DiagnosticsAssayAssay SensitivityDiagnostic Rna Assay ValidationHighly Active Antiretroviral TherapyHiv Nucleic Acid TestingHuman Immunodeficiency Virus TestingLaboratory Diagnosis Of HivQuantitative Determination
A new quantitative reverse transcription (RT)-PCR assay for human immunodeficiency virus type 1 (HIV-1) RNA (Abbott LCx HIV RNA Quantitative assay) has been compared with the Organon NucliSens assay on 521 retrospective samples obtained from HIV-1-positive patients monitored during highly active antiretroviral therapy, 79 of whom were assayed also by the Chiron Quantiplex 3.0 system and on characterized panels. The LCx system showed a moderate correlation (r = 0.795) and gave higher results than the NucliSens system on 245 of 327 concordant positive samples, with similar sensitivity. Correlation with Quantiplex system results was higher (r = 0.943). LCx reproducibility was very good; the procedure was simple, well controlled, and rapid (up to 48 results in 7 h). The HIV RNA quantitative assay on the LCx system is suitable for routine use.
23
Belinda Yen‐Lieberman, Donald Brambilla, James W. Bremer et al. · Journal of Clinical Microbiology · 1996 · 161 citations · Full text