In the routine care of newborns, the first method for bilirubin testing should be a skin test. If the skin test result exceeds 200 micromol/L and other analytes are to be determined with a nonchemical photometric device, then bilirubin can be included in this analysis and the result trusted up to 250 micromol/L. If the skin test result exceeds 200 micromol/L and only bilirubin concentrations are needed, then a standard laboratory method is the first choice, to avoid repeated blood sampling. Bilirubin concentrations from nonchemical photometric devices that exceed 250 micromol/L should be confirmed with standard laboratory methods.