Concepedia

Publication | Closed Access

Transcutaneous Bilirubin Measurement: A Multicenter Evaluation of a New Device

261

Citations

23

References

2001

Year

Abstract

Because the correlation coefficient for HPLC-B and BCF is very similar to that found for HPLC-B and laboratory TSB, BC could be used not only as a screening device but also as a reliable substitute of TSB determination. At higher levels of TSB, in which phototherapy and/or exchange transfusion might be considered, BC performed slightly better than the laboratory. The accuracy and precision of the TcB measurement in this study was observed to be comparable to the standard of care laboratory test.

References

YearCitations

Page 1