IEEE Transactions on Medical Imaging · 2006 · 55 citations · 35 references
EngineeringBiomedical EngineeringRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyComputed TomographyComputed Tomography PhysicsImage ReconstructionMedical ImagingMedical Imaging PhysicsArtifact ReductionBackprojectionCt Image ReconstructionData ConsistencyFiltered Back ProjectionNumerical SimulationSpatial Data ConsistencyTranslational Motion
A basic assumption in the classic computed tomography (CT) theory is that an object remains stationary in an entire scan. In biomedical CT/micro-CT, this assumption is often violated. To produce high-resolution images, such as for our recently proposed clinical micro-CT (CMCT) prototype, it is desirable to develop a precise motion estimation and image reconstruction scheme. In this paper, we first extend the Helgason-Ludwig consistency condition (HLCC) from parallel-beam to fan-beam geometry when an object is subject to a translation. Then, we propose a novel method to estimate the motion parameters only from sinograms based on the HLCC. To reconstruct the moving object, we formulate two generalized fan-beam reconstruction methods, which are in filtered backprojection and backprojection filtering formats, respectively. Furthermore, we present numerical simulation results to show that our approach is accurate and robust.
35
The Mathematics of Computerized Tomography
Frank Natterer, Ge Wang · Medical Physics · 2002 · 2.6K citations · Full text