Effect of pulsed dye-laser wavelength stabilization on spectral overlap in atmospheric NO fluorescence studies

Clayton S. Cooper, Normand M. Laurendeau

Applied Optics · 1997 · 26 citations · 7 references

Concepts

Abstract

We introduce an inexpensive application of a Fabry-Perot etalon to control long-term UV-laser line drift in atmospheric NO laser-induced fluorescence (LIF) measurements by monitoring the visible fundamental of a pulsed dye laser. A linear image sensor captures the interference pattern, and the dye grating can be adjusted to maintain a fixed wavelength through an interface with labview software. Results indicate that the laser wavelength can be fixed to an accuracy of +/-0.0001 nm in the dye fundamental and +/-0.00003 nm in the UV beam. Hence the average error in the LIF signal owing to fluctuations in spectral overlap between the laser and the NO absorption transition decreases from ~5 to ~0.05%, which results in improved measurement accuracy.