Journal of Analytical Atomic Spectrometry · 1996 · 11 citations · 17 references
Natural SciencesChemistryPhysicsAnalytical ChemistryElemental AnalysisSpectroscopyAtomic Emission SpectroscopyAtomic SpectroscopyBackground ReductionDetector SystemsDetector TechnologyAtomic Emission SpectrometryCharge-coupled Device DetectionDetection Limit AssessmentInductively Coupled Plasma Atomic Emission SpectroscopyInstrument Detection LimitMethod Detection LimitSignal-to-background OptimizationSignal-to-background Ratio
The use of the signal-to-root background ratio (SRBR) of a spectral line as the measurable parameter chosen as the criterion for single-element optimization of an inductively coupled plasma with charge-coupled device detection is described. The theoretical background to the choice of the SRBR, rather than the more usual signal-to-background ratio (SBR), is given. Single-element optimization of the carrier gas flow rate and viewing height using both atomic and ionic lines from ten elements is described. The improvement in the detection limit by using the SRBR over the SBR varies by a factor of 1.0–4.8. For example, the detection limit for the Mn II emission line at 257.610 nm is improved from 13.4 ng ml–1 by SBR optimization to 2.8 ng ml–1 by maximizing the SRBR.
17