IEEE Transactions on Nuclear Science · 2008 · 12 citations · 14 references
EngineeringElectrical EngineeringOptoelectronicsRadiation DetectionRadiation MeasurementDetector TechnologyRadiation DetectorsSemiconductor DetectorsDetector BiasingDetector Energy ResolutionDetector Saturation EffectsPolarizationRadiation Energy Resolution
Pulsed bias shutdown operation, in which the bias voltage supply is periodically shut down for a certain time, is suitable for suppressing the polarization effect in a CdTe radiation detector. The duration of the bias voltage shutdown should be short to decrease the dead time and ensure the continuity of the measurement; however, the polarization effect may not be suppressed if the duration is too short. Therefore, we investigated bias voltage shutdown durations and intervals that stabilize the detector performance. A stacked type detector consisting of four 1-mm-thick CdTe diodes was used. At 35degC, the energy resolution for the 511 keV photopeak of the <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">22</sup> Na source saturates at around 30 min after the measurement start and then does not change up to 300 min when the bias voltage is shut down for 10 ms every 5 min. Similarly at 60 <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">deg</sup> C, the energy resolution for the 511 keV photopeak is unchanged for 300 min when the bias voltage is shut down for 10 ms every 10 s.
14
Polarization in Cadmium Telluride Nuclear Radiation Detectors