Development of an orthogonal-stripe CdZnTe gamma radiation imaging spectrometer
23
Citations
12
References
1996
Year
Millimeter Wave TechnologyEngineeringRadiation DetectionPhysicsSpatial Resolution CapabilitiesHealth SciencesSpectroscopyAnalytical InstrumentationApplied PhysicsStripe PitchInstrumentationSynchrotron RadiationRadiation ImagingRadio TelescopeOptoelectronicsRadiologyAstronomical Instruments
We report performance measurements of a sub-millimeter resolution CdZnTe strip detector developed as a prototype for astronomical instruments operating with good efficiency in the 30-300 keV photon energy range. The prototype is a 1.4 mm thick, 64/spl times/64 contact stripe CdZnTe array of 0.375 mm pitch in both dimensions. Pulse height spectra were recorded in orthogonal-stripe coincidence mode which demonstrate room-temperature energy resolution <10 keV (FWHM) for 122 keV photons with a peak-to-valley ratio >5:1. Good response is also demonstrated at higher energies using a coplanar grid readout configuration. Spatial resolution capabilities finer than the stripe pitch are demonstrated. We present the image of a /sup 133/Ba source viewed through a collimator slit produced by a 4/spl times/4 stripe detector segment. Charge signals from electron and hole collecting contacts are also discussed.