British Journal of Radiology · 1953 · 22 citations · 3 references
MedicineRadiation OncologyRadiologyRadiation DosimetryRadiation MeasurementRadiation SafetyExternal Beam DosimetryExternal Beam RadiotherapyIn Vivo DosimetryPhantom DosimetryRadiation Dose MeasurementRadiation TherapyBeta DosimetryBeta Radiation DosimetryCollimator DesignDepth Dose DistributionOrgan DosimetryPercentage Depth DoseSurface Dose MeasurementTherapeutic Dosimetry
A clinical β-ray applicator using cerium 144 is described. When the source is 14·5 mc, the dose-rate on the skin is 59 r/min. Measurements show that depth doses obtained with this applicator are an improvement over those obtained with phosphorus 32, strontium 90 and radium β-ray applicators. The effect of source-skin distance, collimator design, filtration and backscatterers upon dose-rate and percentage depth dose have been investigated. As a result, a form of “β-ray teletherapy” unit is proposed using a 100 mc source of cerium 144 at a source-skin distance of 3·4 cm. Such a unit would give a surface dose-rate of 94 r/min and an intensity of half this value at a depth of 3·2 mm.
3
Dose Measurement in Beta-ray Therapy
G. W. Blomfield, F. W. Spiers · British Journal of Radiology · 1946 · 17 citations