Journal of Analytical Atomic Spectrometry · 1995 · 22 citations · 3 references
Natural SciencesChemistryAnalytical ChemistryElemental AnalysisSpectroscopyAbsorption SpectroscopyAtomic SpectroscopyHeavy Metal AnalysisInstrumental AnalysisLimit Of DetectionTrace Element AnalysisAnalytical Detection LimitsAnalytical Detection Of Heavy MetalsBlood Cadmium LevelsElectrothermal Atomic Absorption SpectrometryHeavy Metal DetectionPhotometric Error AnalysisPhotometric MeasurementQuantitative Analysis Of Elements
A theoretical and experimental study was made of the effect of photometric measurement error on the detection limits in Zeeman-effect electrothermal atomic absorption spectrometry. The results were used to develop a simple method for calculating the detection limits using four experimentally determined parameters, namely the baseline offset compensation time, the integration time, the light flux ‘energy’ and the background absorbance. A comparison was made of the calculated and measured detection limits for a number of elements in pure water solutions, and also in NaCl, blood and sea-water matrices. The calculated detection limits were found to be in agreement with the experimental data for all the elements studied. The performance of the method was demonstrated by optimizing the conditions for the determination of cadmium in blood.
3