Optics Express · 2007 · 152 citations · 22 references
EngineeringElectrical EngineeringRadiation DetectionRadiation MeasurementUncertainty QuantificationCalibrationDetector EfficiencyDetector TechnologyMeasurement UncertaintyOptical DetectorsCalibration TechniquesCounting Efficiency CalibrationDetector CharacterizationDetectorsPhoton CountingPhoton DetectionPhoton DetectorsPhoton-counting Detectors
We have characterized an independent primary standard method to calibrate detection efficiency of photon-counting detectors based on twophoton correlations. We have verified this method and its uncertainty by comparing it to a substitution method using a conventionally calibrated transfer detector tied to a national primary standard detector scale. We obtained a relative standard uncertainty for the correlated-photon method of 0.18 % (k=1) and for the substitution method of 0.17 % (k=1). From a series of measurements we found that the two independent calibration techniques differ by 0.14 (14) %, which is within the established uncertainty of comparison. We believe this is the highest accuracy characterization and independent verification of the correlated-photon method yet achieved.
22
Quantum Fluctuations and Noise in Parametric Processes. I.
W. H. Louisell, A. Yariv, A. E. Siegman · Physical Review · 1961 · 718 citations
Measurement-induced entanglement for excitation stored in remote atomic ensembles
C. W. Chou, Hugues de Riedmatten, D. Felinto et al. · Nature · 2005 · 589 citations · Full text
Optical radiation measurements.
F. Grum, Richard J. Becherer · 1979 · 291 citations