Production of compound A and carbon monoxide in circle systems: an <i>in vitro</i> comparison of two carbon dioxide absorbents*

Michel Struys, M.P. Bouche, G Rolly, Y. D. I. Vandevivere, D. Dyzers, W. Goeteyn, L Versichelen, Jan F. P. Van Bocxlaer, Éric Mortier

Anaesthesia · 2004 · 24 citations · 14 references

Concepts

Abstract

Two new generation carbon dioxide absorbents, DrägerSorb Free and Amsorb Plus, were studied in vitro for formation of compound A or carbon monoxide, during minimal gas flow (500 ml x min(-1)) with sevoflurane or desflurane. Compound A was assessed by gas chromatography/mass spectrometry and carbon monoxide with continuous infrared spectrometry. Fresh and dehydrated absorbents were studied. Mean (SD) time till exhaustion (inspiratory carbon dioxide concentration >or= 1 kPa) with fresh absorbents was longer with DrägerSorb Free (1233 (55) min) than with Amsorb Plus (1025 (55) min; p < 0.01). For both absorbents, values of compound A were < 1 ppm and therefore below clinically significant levels, but were up to 0.25 ppm higher with DrägerSorb Free than with Amsorb Plus. Using dehydrated absorbents, values of compound A were about 50% lower than with fresh absorbents and were identical for DrägerSorb Free and Amsorb Plus. With dehydrated absorbents, no detectable carbon monoxide was found with desflurane.