Medical Physics · 2009 · 27 citations · 13 references
MedicineRadiation OncologyRadiologyRadiation DosimetryRadiation MeasurementTreatment PlanningIn Vivo DosimetryIonization Chamber DosimetryPatient-specific DosimetryPhantom DosimetryRadiation Dose MeasurementRadiation TherapyGafchromic Film DosimetryHelical TomotherapyLongitudinal Case StudyLongitudinal Correlation StudyPhantom ValidationPhantom-based Dose MeasurementRadiation Dose AssessmentTherapeutic Dosimetry
PURPOSE: The Caribbean Radiation Oncology Center acquired a DELTA4 diode phantom for helical tomotherapy IMRT QA and presents the results of their first 264 clinical cases. METHODS: The validation consisted of several case studies comparing existing ionization chamber and Gafchromic film IMRT QA results to diode phantom results, along with a longitudinal study analyzing the IMRT QA results against other machine QA procedures for a complete sample of IMRT patients. RESULTS: The case studies resulted in a maximum observed difference of 0.7% between the diode phantom and the ionization chamber measurements in low dose-gradient regions. Over the longitudinal study, every IMRT QA plan passed a gamma specification of a 3%/3 mm and 98% of the diodes yielded a value of less than 1. In addition, the mean 90% isodose absolute difference for all plans was 0.05% with a (lsigma) standard deviation of 1.19%. CONCLUSIONS: The phantom measurements closely match the planned dose distributions in high and low dose-gradient regions. In addition, a significant positive statistical correlation was determined between the IMRT QA, daily QA, and rotational variation output measurements. Together, these results signify high degree of accuracy of both the DELTA4 phantom as well as the TomoTherapy Hi-Art system.
13
A technique for the quantitative evaluation of dose distributions
Daniel A. Low, William B. Harms, Sasa Mutic et al. · Medical Physics · 1998 · 2.9K citations · Full text
Tomotherapy: A new concept for the delivery of dynamic conformal radiotherapy
T Mackie, Timothy Holmes, Stuart Swerdloff et al. · Medical Physics · 1993 · 1.1K citations · Full text
Gary A. Ezzell, James M. Galvin, Daniel A. Low et al. · Medical Physics · 2003 · 877 citations · Full text