Abstract X‐ray spectroscopy has been widely used for chemical analysis since its discovery in 1895. X‐ray fluorescence (XRF) has been used for over 60 years for identifying the elemental composition in most kinds of samples and in combination with modern semiconductor detector technology XRF has developed into a versatile, multi‐element method. With the high X‐ray flux attainable from the synchrotron source, new spectroscopic techniques have become available, such as extended x‐ray absorption fine structure (EXAFS) and x‐ray absorption near‐edge spectroscopy (XANES), for studying the chemical states of the different elements. With the use of high‐intensity x‐ray microbeams all of these different analytical methods can be performed on a microscopic level. By scanning the microbeam over a sample, the distribution of different elements can be determined. A description of the basic principles of x‐ray microbeam spectroscopy, and the historical background, is given and comparisons with other techniques are discussed. Different techniques for focusing of x‐rays, detection and image reconstruction are described. Examples of applications and future perspectives are discussed with particular emphasis on capillary optical systems.