Journal of the Optical Society of America B · 1985 · 173 citations · 12 references
Natural SciencesChemistryAnalytical ChemistrySpectroscopic AnalysisSpectroscopyFourier Transform SpectroscopyInfrared InstrumentationInfrared SpectroscopyPrecision SpectroscopyRotational SpectroscopyCalibration Of Far-infrared InstrumentsFar-infrared InstrumentationFar-infrared ObservationsFar-infrared SpectroscopyFourier Transform Infrared SpectroscopyHigh-resolution Infrared SpectroscopyInfrared SpectraSpectrometer Resolution
Far-infrared pure rotational spectra of H216O, H218O, HD16O, HD18O, D216O, and D218O have been measured with a high-resolution Fourier-transform spectrometer with an apparatus function of 0.004 cm−1 in the region from 20 to 350 cm−1. Altogether about 1200 lines have been assigned and measured with an accuracy that is believed to be ~0.0002 cm−1 (~6 MHz), so that many of the frequencies should be of use in the calibration of far-infrared spectrometers.