2002 · 19 citations · 9 references
EngineeringElectrical EngineeringHeat TransferInstrumentationThermal AnalysisFrequency MetrologyOscillatorsCesium Frequency StandardsFrequency StabilityFrequency Stability AnalysisFrequency StandardsFrequency SynthesisFrequency SynthesizersSpectral Purity AssessmentTemperature Effects On Spectra
This paper reviews the environmental effects in synthesizers designed to support a frequency stability of 10/sup -13/ /spl tau//sup -1/2/ in short term and 10/sup -17/ in the long term. Specifically we consider the effects of temperature, pulling by spurious spectral lines, vibration effects, and pickup of spurious rf signals. We show that the temperature coefficient of the new NIST HR1 synthesizers is less than 1 ps/K and that the pulling from spectral purity is less than 3/spl times/10/sup -20/ in NIST-7, our primary thermal cesium beam standard. The pulling for slow cesium standards should be lower. We also show that the pulling due to spurious lines in Ramsey standards with narrow line widths can be manipulated to examine spectral pulling. The fractional frequency stability is better than 3/spl times/10/sup -14/ /spl tau//sup -1/2/ for measurement times out to 10/sup 4/ s and reaches 10/sup -16/ in about 15 minutes in a standard laboratory environment of roughly +/-0.5 K without the need of additional thermal regulation.
9
LOCAL OSCILLATOR INDUCED INSTABILITIES IN TRAPPED ION FREQUENCY STANDARDS
G.J. Dick · 1987 · 182 citations
A cesium fountain frequency standard: preliminary results
A. Clairon, Philippe Laurent, G. Santarelli et al. · IEEE Transactions on Instrumentation and Measurement · 1995 · 165 citations
Natural Sciences, Physics, Optics +12