Application of an Electrochemical Immunosensor with a MWCNT/PDAA Modified Electrode for Detection of Serum Trypsin

Qiang Yi, Qicai Liu, Feng Gao, Qingquan Chen, Guina Wang

Sensors · 2014 · 23 citations · 17 references

DOIFull text

Open access

Concepts

Abstract

OBJECTIVE: To establish an electrochemical immunosensor for the determination of serum trypsin levels using a multiwall carbon nanotubes (MWCNTs)-composite-modified electrode. METHOD: A MWCNT composite coated on the surface of bare gold electrodes was used for fixation of an anti-trypsin antibody. The assembly process and the performance indicators, including sensitivity, linear range of detection, anti-jamming performance, and stability, of the electrochemical immunosensor were examined by cyclic voltammetry (CV) and electrochemical impedance spectroscopy (EIS). RESULTS: With optimized experimental conditions, the difference of the current value measured by differential pulse voltammetry (DPV) showed a linear relationship with the concentration of serum trypsin within 0.10-100 ng/mL. The detection limit for trypsin using this sensor was 0.002 ng/mL. CONCLUSIONS: The electrochemical immunosensor built using the MWCNT-composite-modified electrode is simple to operate and has a fast response time, along with a wide linear range, high sensitivity, and accuracy, making it suitable for serum trypsin detection.

References

17