In Vitro Performance Testing of the Novel Medspray® Wet Aerosol Inhaler Based on the Principle of Rayleigh Break-up

Anne H. de Boer, Jeroen Wissink, Paul Hagedoorn, Iwan Heskamp, Wilbur de Kruijf, Ralf Bünder, Pieter Zanen, Paul Munnik, C.J.M. van Rijn, Henderik W. Frijlink

Pharmaceutical Research · 2007 · 17 citations · 16 references

DOIFull text

Open access

Concepts

Abstract

PURPOSE: A new inhaler (Medspray) for pulmonary drug delivery based on the principle of Rayleigh break-up has been tested with three different spray nozzles (1.5; 2.0 and 2.5 mum) using aqueous 0.1% (w/w) salbutamol and 0.9% (w/w) sodium chloride solutions. MATERIALS AND METHODS: Particle size distributions in the aerosol were measured with the principles of time of flight (APS) and laser diffraction (LDA). RESULTS: The Medspray inhaler exhibits a highly constant droplet size distribution in the aerosol during dose emission. Droplets on the basis of Rayleigh break-up theory are monodisperse, but due to some coalescence the aerosols from the Medspray inhaler are slightly polydisperse. Mass median aerodynamic diameters at 60 l.min(-1) from APS are 1.42; 1.32 and 1.27 times the theoretical droplet diameters (TD's) and median laser diffraction diameters are 1.29; 1.14 and 1.05 times TD for 1.5; 2.0 and 2.5 mum nozzles (TD: 2.84; 3.78 and 4.73 mum respectively). CONCLUSIONS: The narrow particle size distribution in the aerosol from the Medspray is highly reproducible for the range of flow rates from 30 to 60 l.min(-1). The mass median aerodynamic droplet diameter can be well controlled within the size range from 4 to 6 mum at 60 l.min(-1).

References

16