Applied Optics · 1998 · 123 citations · 10 references
EngineeringBiomedical EngineeringElectrical EngineeringBiomedical OpticsPhotonicsSensorsOptical DetectorsOptical InstrumentationOptical MetrologyOptical SensingOptical SensorsTime MeasurementLateral ResolutionLong-term StabilityOptical Measurement SystemsPhoton CountingRepeatabilitySingle-photon DetectionStand-off DistanceTime-of-flight Measurement
The design and implementation of a prototype time-of-flight optical ranging system based on the time-correlated single-photon-counting technique are described. The sensor is characterized in terms of its longitudinal and transverse spatial resolution, single-point measurement time, and long-term stability. The system has been operated at stand-off distances of 0.5-5 m, has a depth repeatability of <30 mum, and has a lateral spatial resolution of <500 mum.
10
Avalanche photodiodes and quenching circuits for single-photon detection
S. Cova, Massimo Ghioni, Andrea L. Lacaita et al. · Applied Optics · 1996 · 1.1K citations
Active, optical range imaging sensors
Paul J. Besl · Machine Vision and Applications · 1988 · 554 citations
20-ps timing resolution with single-photon avalanche diodes
S. Cova, Andrea L. Lacaita, Massimo Ghioni et al. · Review of Scientific Instruments · 1989 · 137 citations
Laser depth measurement based on time-correlated single-photon counting
J. S. Massa, Andrew Wallace, Gerald S. Buller et al. · Optics Letters · 1997 · 101 citations