IEEE Sensors Journal · 2014 · 19 citations · 23 references
EngineeringBioengineeringBiomedical EngineeringInstrumentationSensor FusionSensorsChemical SensorsFault DetectionPh AnalysisSensor DiagnosticsSmart SensorsData FusionMiniaturized Reference ElectrodesPh MeasurementPh MonitoringPh SensingPh SensorsSensingSensor Fault Diagnosis
This paper concerns the data fusion (DF) and fault diagnosis (FD), which could increase the reliability of measured results when one sensing window was faulty. The flexible arrayed pH sensor was fabricated by radio frequency sputtering system and screen-printed technology, which the ruthenium dioxide was deposited on the polyethylene terephthalate substrate as sensitive membranes and the miniature reference electrodes were fabricated by screen-printed technology. We measured pH buffer solutions for five times and the flexible arrayed pH sensor had a wide sensing range of pH 1-13 solutions. The average sensitivity and linearity were 47.70 mV/pH and 0.839, respectively. We used the Laboratory Virtual Instrumentation Engineering Workbench to do deficient diagnosis which could remove the fault sensing window. The average DF with FD, self-adaptive data fusion with FD and coefficient of variance data fusion with FD could increase sensitivity and linearity ~ 22% and 0.14, respectively.
23
A flexible pH sensor based on the iridium oxide sensing film
Wen-Ding Huang, Hung Cao, Sanchali Deb et al. · Sensors and Actuators A Physical · 2011 · 358 citations