Clinical Chemistry · 2014 · 128 citations · 17 references
EngineeringHematologyAnalytical ChemistryClinical BiochemistryGood Laboratory PracticeBioanalytical Method ValidationClinical HematologyGlycemic MonitoringPoint-of-care Diagnostic DevicesPoint-of-care DiagnosticsAssayBlood Group Reference TestingHemoglobin A1c TestingHemoglobin LevelsHemoglobin-based EstimationLaboratory Testing StandardsReference Laboratory StandardsReference Measurement Procedures
BACKGROUND: In 2009, we investigated the conformance of 8 hemoglobin A(1c) (Hb A(1c)) point-of-care (POC) instruments. Since then, instruments have improved and new devices are available on the market. In this second study, we evaluated the performance of DCA Vantage, Afinion, InnovaStar, Quo-Lab, Quo-Test, Cobas B101, and B-analyst Hb A(1c) POC instruments. METHODS: Clinical and Laboratory Standards Institute protocols EP-5 and EP-9 were applied to investigate imprecision, accuracy, and bias. We assessed bias using the mean of 3 certified secondary reference measurement procedures (SRMPs). Assay conformance with the National Glycohemoglobin Standardization Program (NGSP) certification criteria was also evaluated. Interference of common Hb variants was investigated for methods that could work with hemolysed material. RESULTS: The total CVs for all instruments, except for the DCA Vantage at a high Hb A(1c) value, were ≤3.1% in SI units and ≤2.1% in Diabetes Control and Complications Trial (DCCT) units. Afinion, DCA Vantage, B-analyst, and Cobas B101 instruments passed the NGSP criteria with 2 different reagent lot numbers. Quo-Test, Quo-Lab, and InnovaStar instruments had a negative bias compared to the mean of the 3 SRMPs and failed NGSP criteria. Most of the common Hb variants did not interfere with the investigated instruments, except Hb AE for the Cobas B101. CONCLUSIONS: Afinion, DCA Vantage, Cobas B101, and B-analyst instruments met the generally accepted performance criteria for Hb A(1c). Quo-Test, Quo-Lab, and InnovaStar met the criteria for precision but not for bias. Proficiency testing should be mandated for users of Hb A1c POC assays to ensure quality.
17
R. C. Turner, Charles Fox, Matthews et al. · The Lancet · 1998 · 8.4K citations
Standards of Medical Care in Diabetes—2013
Diabetes Care · 2012 · 4.4K citations · Full text
Standards of Medical Care in Diabetes—2011
Diabetes Care · 2010 · 2.9K citations · Full text