IEEE Transactions on Medical Imaging · 1985 · 16 citations · 5 references
MedicineNuclear MedicineRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyCardiac ImagingCardiovascular ImagingComputed TomographyComputed Tomography PhysicsImage ReconstructionMedical ImagingCardiac Computed TomographyComputed Tomography ReconstructionCt Image ReconstructionRadioactivity MeasurementReconstruction AlgorithmsTomographic ImagingTomographic Imaging SystemsTomographic Reconstruction
Cardiac emission tomography using a seven-pinhole collimator has received only little appreciation as a diagnostic imaging technique. The main reasons are the limited angular sampling of the seven-pinhole device and the difficulties encountered in properly positioning the patient relative to the collimator/camera system. In order to overcome these problems, we have developed a modified ART3 algorithm for reconstruction of the radioactivity distribution in the heart. The method is very appropriate for seven-pinhole tomography, as demonstrated by the quality of the reconstructions, by the excellent point source resolution of the system response, and by a comparison to two other suitable reconstruction techniques, viz., SMART and SIRT.
5
Seven-pinhole tomography--a technical description.
Michael T. LeFree, Robert A. Vogel, Dennis Kirch et al. · PubMed · 1981 · 37 citations