IEEE Transactions on Biomedical Engineering · 1999 · 125 citations · 18 references
EngineeringBiomedical EngineeringElectrical EngineeringBiomedical ImagingComputational ElectromagneticsDiagnostic ImagingElectrical CharacterizationElectrical MeasurementImage ReconstructionMedical ImagingMedical Imaging InstrumentationConductivityConductivity ImagingElectrical ImpedanceElectrical Impedance TomographyElectrode Configuration DesignExperimental Tank TestingImpedance ReconstructionTomographic Imaging
A three-dimensional reconstruction algorithm in electrical impedance imaging is presented for determining the conductivity distribution beneath the surface of a medium, given surface voltage data measured on a rectangular array of electrodes. Such an electrode configuration may be desirable for using electrical impedence tomography to detect tumors in the human breast. The algorithm is based on linearizing the conductivity about a constant value. Here, we describe a simple implementation of the algorithm on a four-electrode--by-four-electrode array and the reconstructions obtained from numerical and experimental tank data. The results demonstrate significantly better spatial resolution in the plane of the electrodes than with respect to depth.
18
Electrical Impedance Tomography
Margaret Cheney, David Isaacson, J.C. Newell · SIAM Review · 1999 · 1.4K citations