IEEE Transactions on Magnetics · 1991 · 32 citations · 5 references
EngineeringBiomedical EngineeringFinite Element AnalysisFinite Element MethodsOptimizationMagnetic Circuit AnalysisMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsShielding DesignSuperconducting Magnet SystemsCoil DesignMagnet Design OptimizationMagnetic Coil DesignMagnetic ShieldingMultiobjective Goal ProgrammingNonlinear Least SquaresOptimal DesignSuperconducting Magnet Design
An optical design is presented of a highly homogeneous superconducting coil system with magnetic shielding for magnetic resonance imaging (MRI). The optimal design method is a combination of the hybrid finite-element and boundary-element method for analysis of an axially symmetric nonlinear open-boundary magnetic field problem and a mathematical programming method for solving the corresponding optimization problem. The multiobjective goal programming method and the nonlinear least-squares method are adopted. The optimal design results of 1.5- and 4.7-T-magnet systems with different types of magnetic shielding for whole-body imaging are compared, and the advantages of a combination of active and yoke shields are shown.
5
Considerations in the design of MRI magnets with reduced stray fields
D. Hawksworth, I. McDougall, John Bird et al. · IEEE Transactions on Magnetics · 1987 · 36 citations
Optimal design of MRI magnets with magnetic shielding
Atsushi Ishiyama, M. Hondoh, Nobuyuki Ishida et al. · IEEE Transactions on Magnetics · 1989 · 21 citations