Three‐Dimensional Thoracic Modeling for an Anatomical Compatibility Study of the Implantable Total Artificial Heart

14

Citations

4

References

1999

Year

Abstract

Anatomical compatibility with the thoracic cavity is important in developing a completely implantable total artificial heart (TAH). The purpose of this study was to examine the optimal morphology of our TAH and to develop a computer implantation simulation environment for preoperational discussion. As the first stage, we constructed a prototype simulation system by creating a 3‐D surface model of a thoracic cavity and one of a TAH. In this system, the thoracic surface model, composed of the aorta, pulmonary artery, ventricle resected heart, diaphragm, and thoracic wall, was created based on the organ contours extracted from electrocardiogram (ECG)‐synchronized ultrafast computer tomographic images. The accuracy of the thoracic model was discussed using a phantom. The fitting study can be performed on the computer with the model of the thorax and that of the TAH. In the future, construction of a database for the thoracic cavities of heart failure patients is planned to improve the morphology of the TAH.