Publication | Open Access
Mechanism of Interference of a Polymerized Hemoglobin Blood Substitute in an Alkaline Phosphatase Method
21
Citations
8
References
2000
Year
The interference of polymerized hemoglobin in the alkaline phosphatase assay is a result of an absorbance offset caused by alkali denaturation of hemoglobin. The interference can be corrected or avoided by modifying the calculation or the analytical wavelength. The correction strategy may also be applicable to improving the hemolysis index for this method.