MedicineBiomedical EngineeringRadiologyDiagnostic ImagingUltrasonographyUltrasoundDiagnostic UltrasonographyMedical ImagingMedical Imaging InstrumentationUltrasound ImagingUltrasound InstrumentationUltrasound PhysicsEnsemble SizeOff-line AnalysisUltrasonic ImagingUltrasound Computed TomographyUltrasound DiagnosticsUltrasound Imaging System DesignUltrasound TomographyUser Interface Integration
Under a contract with the National Cancer Institute, we have developed a research interface to an ultrasound system. This ultrasound research interface (URI) is an optional feature providing several basic capabilities not normally available on a clinical scanner. The URI can store high-quality beamformed radio-frequency data to file for off-line processing. Also, through an integrated user interface, the user is provided additional control over the B-mode receive aperture and color flow ensemble size. A third major capability is the ability to record and playback macro files. In this paper, we describe the URI and illustrate its use on three research examples: elastography, computed tomography, and spatial compounding.