IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control · 1986 · 256 citations · 11 references
MedicineBiomedical EngineeringBiomedical ImagingUltrasonographyUltrasoundDiagnostic UltrasonographyUltrasonicsUltrasound ImagingUltrasound InstrumentationUltrasound PhysicsFocus Error AnalysisSpeckle Noise ReductionSpeckle PhenomenaSpeckle ReductionTransducer Array DesignTransducer DesignUltrasonic ImagingUltrasonic Transducer Arrays
In order to design improved spatial compounding ultra- sound scanning systems it is necessary to determine the correlation of speckle patterns as a function of aperture translation. We have con- ducted experiments measuring the speckle correlation with lateral ap- erture translation for linear phased array pulse-echo ultrasonic im- aging systems. Results are presented for variable frequency, range, transducer length, focus error, and reflecting material. Tests were con- ducted on two commercially availahle and one research imaging sys- tem. The measured rates of correlation coefficient decrease are inde- pendent of frequency, reflecting material, and target range when the target is in the focal zone. The experimentally determined correlations are used to derive the optimal spatial separation of images for speckle reduction.
11
Statistics of Speckle in Ultrasound B-Scans
Robert F. Wagner, S.W. Smith, John M. Sandrik et al. · IEEE Transactions on Sonics and Ultrasonics · 1983 · 1.4K citations
Modern Probability Theory and Its Applications.
N. L. J., Emanuel Parzen · The Incorporated Statistician · 1960 · 629 citations