Radiation Protection Dosimetry · 2012 · 12 citations · 7 references
MedicineNuclear EngineeringRadiologyRadiation DosimetryRadiation MeasurementRadiation PhysicsAbsorbed DoseDose CalibrationMonte Carlo DosimetryPhantom DosimetryPhoton DosimetryRadiation Dose MeasurementAbsorbed Dose UncertaintyEquivalent Dose DeterminationGraphite CalorimetryIonizing Radiation DosimetryRadiation Dose AssessmentRadiation Dose EstimationRadiation Source CalibrationWater Phantom Measurements
The ARPANSA calibration service for (60)Co gamma rays is based on a primary standard graphite calorimeter that measures absorbed dose to graphite. Measurements with the calorimeter are converted to the absorbed dose to water using the calculation of the ratio of the absorbed dose in the calorimeter to the absorbed dose in a water phantom. ARPANSA has recently changed the basis of this calculation from a photon fluence scaling method to a direct Monte Carlo (MC) calculation. The MC conversion uses an EGSnrc model of the cobalt source that has been validated against water tank and graphite phantom measurements, a step that is required to quantify uncertainties in the underlying interaction coefficients in the MC code. A comparison with the Bureau International des Poids et Mesures (BIPM) as part of the key comparison BIPM.RI(I)-K4 showed an agreement of 0.9973 (53).
7
Monte Carlo simulation of a typical 60Co therapy source
G. Mora, D. W. O. Rogers · Medical Physics · 1999 · 196 citations · Full text