Physics in Medicine and Biology · 1997 · 25 citations · 12 references
MedicineMaterials ScienceRadiation DosimetryRadiation MeasurementRadiation PhysicsAbsorbed DoseDosimetry InstrumentationIonization Chamber DosimetryPhoton DosimetryRadiation Dose MeasurementReference DosimetryAbsorbed Dose-to-water DeterminationDosimeter CalibrationEquivalent Dose DeterminationIon Beam DosimetryIonizing Radiation DosimetryReference Beam DosimetryReference Value Validation
To derive the absorbed dose to water from a standard of absorbed dose to graphite, the metrology laboratories which apply such a method usually make use of cavity ionization chambers as transfer instruments. In addition, the BNM-LPRI has tested, as such instruments, two types of Fricke dosimeter in its cobalt-60 beam. The two procedures are compared and their results are found to be in good agreement (the difference is less than 0.1%). Both procedures are then taken into account for the calculation of the reference value of absorbed dose to water.
12
A General Theory of Cavity Ionisation
T. E. Burlin · British Journal of Radiology · 1966 · 192 citations